Trans-Data M-2000 Meter Error Codes
ERROR CODE | DESCRIPTION | SEE PARAMETER |
1 | Invalid parameter number for Calibration Mode | Calibration Mode |
2 | Invalid graduation size | 3 |
3 | Invalid decimal position | 2 |
4 | Parameter value must be-On (1) or Off (0) | 22 |
5 | Push to Zero Window must be 0-99 | 21 |
8 | Only 1 will reset Span Table | 15 |
9 | Span exceeds maximum capacity/or entry too small | 15 |
10 | IZSM value can be-On (1) or Off (0) | 45 |
12 | Motion settle time out of range (1-50) | 24 |
13 | Power on units may only be-lb. (0) / kg (1) | 6 |
14 | Invalid time entry HH.MM.SS | 80 |
15 | Invalid date entry YY.MM.DD | 81 |
16 | Motion value is out of range (1-10) | 5 |
17 | Press tare to increment span table; any other key invalid | 16 |
18 | Press tare to decrement span table; any other key invalid | 17 |
19 | Span table cannot be incremented past 1 | 16 |
20 | Parameter memory write error, indicator needs service | Call for Support |
21 | Parameter check sum error, parameters have been lost | Call for Support |
22 | Program check fault; indicator needs service | Call for Support |
23 | Invalid serial port speed setting | 30, 32 |
24 | Invalid serial port parity parameter | 27, 31, 33 |
25 | Cannot increment span table any further | 16 |
26 | Entered offset larger than capacity | 25 |
27 | Invalid string mode for port | 34 & 35 |
28 | Power on zero warning-OFF (0) / ON (1) | 9 |
29 | Channel enable-OFF (0) / ON (1) | 98 |
30 | Only 1 will reset the dead load | 12 |
31 | Sound volume can be between 0-3 | 28 |
32 | Key press feedback can be-OFF (0) / ON (1) | 29 |
33 | Invalid COM string mode parameter | 35 |
34 | Invalid COM port interface value | 47 |
37 | Channel 1 cannot be disabled | 98 |
38 | Invalid Print Select Function Number | Normal Weighing Mode |
40 | Scale channel is not enabled | Normal Weighing Mode |
41 | Push button tare is invalid (Over, Motion, or disabled) | Normal Weighing Mode |
42 | Keyboard tare available on Channel 1 only | Normal Weighing Mode |
43 | Tare greater than capacity | Normal Weighing Mode |
45 | Password – Enable (1)/Disable (0) | 96 |
46 | Invalid entry (Parameter 26–Values of 0,1 & 80 are acceptable) | 26 |
47 | Bad Clock Mode (0-2) | 83 |
48 | Invalid Main Filter Value | 41 |
49 | Invalid Output String Format | 34 |
49 | Invalid Fast-step Threshold Value | 42 |
50 | Zero tracking must be 1-99 or 100, 200, 300 | 23 |
50 | Invalid Fast Step Sensitivity | 43 |
51 | Keypad Tare Function disabled | Tare Scale |
51 | Invalid Fast Step Setting-OFF (0) / ON (1) | 44 |
52 | Tare Function disabled | Tare Scale |
54 | Invalid Test Mode Entry (0-3) | 50 |
56 | Invalid Power Bypass Entry (0 or 1) | 10 |
56 | Invalid Display Update Rate (0-9) | 19 |
56 | Cannot delete ticket; ticket # not found | 89 |
57 | Invalid input range for AD voltage range-default (1) | 11 |
100 | Invalid password # range; can only be 0000-9999 | 96 |
110 | SRAM failure | Call for Support |
120 | Battery or internal power failure | Call for Support |
140 | FLASH memory erase failed | Call for Support |
141 | FLASH memory write failed | Call for Support |
151 | Audit trail CRC failed | Call for Support |
152 | Database CRC failed | Call for Support |
153 | CAL copy CRC failed | Call for Support |
154 | Ticket buffer CRC failed | Call for Support |
185 | SMART wire COM link not responding | 59 |
191 | Channel 1 AD converter not responding | Call for Support |
192 | Channel 2 AD converter not responding | Call for Support |
193 | Channel 3 AD converter not responding | Call for Support |
231 | Ticket does not exist | Ticket Printing |